CHESTGRAPH HI-301

• Two different choices of flow sensors (Turbine and Ultrasonic)
• 7 inch touchscreen monitor
• Rechargeable battery
• 6MWT with optional Pulse Oximeter
• Temperature sensor embedded
• 20,000 test data storage
• Animated guidance for patients
• Direct printing by external USB printer
Details
HI-301 with Turbine Sensor
HI-301 with Ultrasonic Sensor
Specification
Flow
Detection: Ultrasonic Sensor/Turbine Sensor
Range: 0 to +/- 18L/s
Resolution: 0.01L/s
Accuracy: +/- 5% or +/- 0.2L/s, whichever is greater
Volume
Detection: Flow integration
Range: 0 to +/- 10.0L
Resolution: 0.01L
Accuracy: +/- 3% or +/- 0.05L, whichever is greater
LCD
Display: 7inch color touchscreen monitor
Others
Printer: Thermal printer (112mm width)
Data storage: More than 20,000 data
Interface: Wireless communication, USB-A, USB-mini, USB Micro-B
Power: 100-240V 50/60HZ
Dimensions: 230(W) × 240(D) × 85(H)mm
Weight: Approx. 2.5kg
Measurement Items
Spirometry: SVC, FVC, TV, FEV0.5, FEV1, FEV3, FEV6, FEV1/FVC, FEV1/SVC, FEV1/SVC, FEV1/FEV6, MMEF(FEF25-75), PEF, FEF25, FEF50, FEF75, FEF90, FEF50/FEF75 FEF75-85, FEF200-1200, OI, ATI, PEF time, FET, ExtrapV, ExtrapV%, FIVC, FIV0.5, FIV1, FIV1/FVC, FIV1/FIVC, PIF, FIF50, FEF50/FIF50, FIF50/FEF50, CVI, MVV43, FVC+FEV1, MVV, MV, TV, RR, BR, VR, BD test, Bronchial Challenge test, 6MWT

CHEST M.I. INC. was founded in Tokyo in 1965 as Sanwa Seiki Corp., which developed a water-sealed spirometer. The company was re-named CHEST M.I. (Medical Instruments) INC to clarify its specialty in 1974 and began to export into international market. In 1971, they developed the Japan’s first computerized Spirometer, DISCOM. In 1976, the comprehensive pulmonary function testing system, CHESTAC was released.
CHEST have been successively developing new products, such as ASTOGRAPH, airway hyper reactivity test using Forced Oscillation in 1978 (the first of its kind), RHINOGRAPH, a computerized nasal resistance meter in 1980. In 1982, innovative portable, calibration-free spirometer, HI-498 was developed. This product enjoyed big success in Europe and the US, and built-up the company reputation in global market.
APNOMONITOR, sleep apnea monitor at home was developed in 1982, PEAKMAN, the handheld electronic Peakflow meter in 1993 (the first of its kind), HOTMATE, 24hr respiratory condition monitor (the first of its kind) in 1998, APNOMONITOR NEURO, Polysomnograph in 2002. More recently, they have succeeded in developing MostGraph, the colored 3D analyzer of respiratory resistance & reactance in COPD and Asthma (the first of its kind). This new method without patient cooperation is expected to improve the diagnosis of respiratory disease.

jQuery(document).ready(function($){ $(".wd-entities-title a").each(function(){ var text = $(this).html(); consol.log(text); text = text.replace(/ مدل /g, "